Relative time clock on MIN, MAX, and AVG provides a time reference for major events
Electronic Offset function allows compensation of thermocouple errors to maximize overall accuracy
Measures J, K, T, and E-types of thermocouples
Readout in ºC, ºF, or Kelvin (K)
Splash and dust resistant case protected by impact absorbing holster
User-friendly front panel is easy to set up and operate
Sleep mode increases battery life; typical 1000-hour battery life
Battery door allows easy battery replacement without breaking the calibration seal
Optional ToolPak accessory allows the thermometer to hang from any metal object (with the rare earth magnet) or secure around a pipe (with hook-and-loop straps) for hands-free operation.
1 year warranty
The Fluke 53 II B and 54 II B include all the features listed above plus:
Large backlit dual display shows any combination of T1, T2 (52 and 54 only), T1-T2 (52 and 54 only), plus MIN, MAX, or AVG.
Additional thermocouple types R, S, and N (for a total of 7 different types)
Recall function allows logged data to be easily reviewed on the meter display
USB communication port allows data to be exported to optional FlukeView Forms® Temperature PC software for further analysis and graphing
Data Logging up to 500 points of data with user adjustable recording interval
Thông số kỹ thuật Đồng hồ đo nhiệt độ Fluke 54 II
Specifications
Temperature accuracy
Above -100 °C:
J, K, T, E, and N-type: ±[0.05% + 0.3°C]*
R and S-type: ±[0.05% + 0.4°C]*
Below -100 °C:
J, K, E, and N-types: ±[0.20% + 0.3°C]*
T-type: ±[0.50% + 0.3°C]
Temperature
J-type:
-210 °C to 1200 °C
K-type:
-200 °C to 1372 °C
T-type:
-250 °C to 400 °C
E-type:
-150 °C to 1000 °C
N-type:
-200 °C to 1300 °C *
R and S-type:
0 °C to 1767 °C *
Temperature scale
ITS-90
Applicable standards
NIST-175
Display Resolution
0.1 °C, 0.1 K < 1000
1°C, 1 K >=1000
Note
*
Only the Fluke Models 53 II B and 54 II B thermometers are capable of measuring N, R, and S-type thermocouples.
Environmental Specifications
Operating Temperature
-10 °C to 50 °C
Storage Temperature
-40 °C to 60 °C
Humidity (Without Condensation)
0% to 90%; 0 °C to 35 °C
0% to 70%; 0 °C to 50 °C
Safety Specifications
Overvoltage category
CSA C22.2 No. 1010.1 1992; EN 61010 Amendments 1,2
Agency Approvals
CE, CSA, TÜV (pending)
Mechanical & General Specifications
Size
173 mm L x 86 mm W x 38 mm
Weight
400 g
Batteries
3 AA batteries; typical 1000-hour life
TKTECH.VN Cung cấp Đồng hồ đo nhiệt độ Fluke 54 II